LANDTEK DP29-40 Portable Chilled Mirror Dew Point Instrument

Art. number:
Price (BGN):
0.00 лв.
Price (EUR):
€0.00

Description

The LANDTEK DP29-40 is a multifunctional portable dew point instrument based on the chilled mirror measurement principle. It is intended for field and laboratory applications where accurate and stable determination of moisture in gas is required.

According to the datasheet, the DP29-40 is specifically designed for air dew point measurement. It is suitable for use as a reference instrument in electronic component durability testing, for dew point control in firing furnaces, and for dew point control in fuel cell test systems.

The instrument operates by cooling a mirror surface in a controlled manner until condensation begins to form. The temperature at that moment is used to determine the dew point.

The DP29-40 is designed to provide accurate and repeatable measurements, while the system offers stable operation without the need for frequent recalibration.

A built-in electronic sample gas flow monitor helps maintain stable measurement conditions during operation.

The model provides an RS-232 interface for connection to a PC, allowing measurement data to be transferred for further processing.

The LANDTEK DP29-40 is suitable for laboratory measurements, dry-air monitoring, furnace processes, test systems, and other applications requiring reliable air dew point measurement.

 

For more information:
Chilled Mirror Dew Point Instrument DP29-40 - Guangzhou Landtek Instruments Co.,LTD

Specifications

Parameter Value
Model DP29-40
Measurement Principle Chilled mirror
Measurement Range at Temperature Below 0 °C Down to -40 °C
Measurement Range at Temperature Below 10 °C Down to -40 °C
Measurement Range at Temperature Below 20 °C Down to -35 °C
Measurement Range at Temperature Below 35 °C Down to -30 °C
Accuracy ≤ ±0.2 °C, ±1 digit
Repeatability ≤ ±0.1 °C, ±1 digit
Resolution 0.1 °C
Cooling Method Semiconductor refrigeration
Inlet Flow 15–60 L/h, typically 30–40 L/h
Inlet Pressure 10 mbar–10 bar (1 kPa–1 MPa)
Response Rate Up to 2 °C/s
Default Output USB
Optional Output ±10 mV/°C (0 °C = 0 mV) or 4–20 mA
Operating Temperature 0 to +50 °C
Operating Humidity < 90% RH, no condensation
Power Supply 220 VAC ±10%, 50/60 Hz
Power Consumption Approx. 160 W
Dimensions 342 × 140 × 326 mm
Weight Approx. 7.22 kg
Mirror Check Manual check using “CH” mode
Instrument Check Manual check using the Test Cooling function
Standard Accessories Main unit, 3 m FEP inlet tube with connector, operation manual, packaging box
Optional Accessories RS-232C data cable with software

 