The kit includes the PVM-1530 photovoltaic meter, two IRM-1 solar radiation and temperature meters, and the CMP-1015-PV digital clamp meter for photovoltaic systems.
The Sonel PVM-1530 Max is an extended kit for testing, diagnosing and assessing the safety and performance of photovoltaic installations. The PVM-1530 supports Category 1 measurements under EN 62446-1 and Category 2 I‑V characteristic measurements under EN 62446-1 and EN 61829. It is suitable for standard, bifacial and high-efficiency PV modules.
The tester measures I‑V and P‑V characteristics and converts results to Standard Test Conditions (STC) in accordance with EN IEC 60891. The two IRM-1 meters measure solar irradiance, module temperature and ambient temperature. Their data helps assess measurement conditions, support PV system design and identify shading. If communication is temporarily interrupted, the reSYNC function automatically supplements the results with the parameters recorded by the IRM-1 meters once the connection is restored.
The IRM-1 measures solar irradiance in W/m² or BTU/ft²·h, PV module and ambient temperature, module tilt and orientation. Each IRM-1 includes a built-in compass, tilt sensor, LoRa communication and a data logger with memory for up to 5,000 records. Its enclosure is rated IP65.
The CMP-1015-PV measures DC voltage and current in photovoltaic installations up to 1500 V. Its VFD function supports measurements downstream of frequency converters. The meter also features a graphing data logger and a high-brightness display for outdoor use.
The PVM-1530 features touchscreen operation, a bright colour display, structured measurement memory, Bluetooth and Wi-Fi connectivity. Data can be transferred to a computer by cable or wirelessly. Sonel Reader enables users to view, export and print results. Sonel Reports Plus is available as optional software for preparing electric shock protection reports. The kit is part of the Sonel MeasureEffect™ platform.
For more information: Sonel PVM-1530 Max official product page
| Parameter | Measuring range | Display range | Resolution | Accuracy |
|---|---|---|---|---|
| Voltage | ||||
| AC voltage | 0.0…1000.0 V | 0.0…1000.0 V | 0.1 V | ±(2% of measured value + 6 digits) |
| DC voltage | 0.0…1500.0 V | 0.0…1500.0 V | 0.1 V | ±(0.5% of measured value + 2 digits) |
| Frequency | 10.0…100.0 Hz | 0…100.0 Hz | 0.1 Hz | ±(0.5% of measured value + 2 digits) |
| Short-circuit current ISC | ||||
| At 1500 V DC | 0.00…30.00 A | 0.00…30.00 A | 0.01 A | ±(1% of measured value + 2 digits) |
| At 1000 V DC | 0.00…40.00 A | 0.00…40.00 A | 0.01 A | ±(1% of measured value + 2 digits) |
| Insulation resistance | ||||
| AC-side insulation resistance | ||||
| Test voltage 250 V | 250 kΩ…200 MΩ according to EN IEC 61557-2 |
0.0 kΩ…200.0 MΩ | from 0.1 kΩ | ±(3% of measured value + 8 digits) |
| Test voltage 500 V | 500 kΩ…500 MΩ according to EN IEC 61557-2 |
0.0 kΩ…500.0 GΩ | from 0.1 kΩ | ±(3% of measured value + 8 digits) |
| Test voltage 1000 V | 1000 kΩ…1.000 GΩ according to EN IEC 61557-2 |
0.0 kΩ…1.000 GΩ | from 0.1 kΩ | ±(3% of measured value + 8 digits) |
| DC-side insulation resistance | ||||
| Test voltage 250 V / 500 V / 1000 V / 1500 V | 250 kΩ…500 GΩ according to EN IEC 61557-2 |
0.0 kΩ…500.0 GΩ | from 0.1 kΩ | ±(8% of measured value + 8 digits) |
| Resistance of protective conductors and equipotential bonding | ||||
| Measurement with ±200 mA current | 0.10 Ω…1999 Ω according to EN IEC 61557-4 |
0.00 Ω…1999 Ω | from 0.01 Ω | from ±(2% of measured value + 3 digits) |
| Operating current | As for CMP-1015-PV | As for CMP-1015-PV | As for CMP-1015-PV | As for CMP-1015-PV |
| Power measurement | 0.0…999.0 kW | 0.0…999.0 kW | 0.1 kW | ±(6% of measured value + 5 digits) |
| Parameter | Measuring range | Display range | Resolution | Accuracy |
|---|---|---|---|---|
| Solar irradiance | ||||
| Measurement in W/m² | 0…1400 W/m² | 0…1400 W/m² | 1 W/m² | ±(5% of measured value + 2 digits) |
| Measurement in BTU/ft²·h | 0…444 BTU/ft²·h | 0…444 BTU/ft²·h | 1 BTU/ft²·h | ±(5% of measured value + 2 digits) |
| PV module and ambient temperature | ||||
| Measurement in °C | −20.0…100.0 °C | −20.0…100.0 °C | 0.1 °C | ±(1% of measured value + 5 digits) |
| Measurement in °F | −4.0…212.0 °F | 4.0…212.0 °F | 0.1 °F | ±(1% of measured value + 5 digits) |
| Tilt angle | −90°…+90° | −90°…+90° | 1° | ±2° |
| Direction — compass | 0°…360° | 0°…360° | 1° | ±5° |
Measured value — the value recorded by the instrument.
Note: In the supplied specification, the display range for insulation resistance at a 500 V AC test voltage is listed as 500.0 GΩ, while the measuring range is up to 500 MΩ. For temperature in °F, the measuring range starts at −4.0 °F, while the display range starts at 4.0 °F. Values are retained as stated in the source.