LANDTEK SL-1256DOS Logging Personal Noise Dose Meter

LANDTEK SL-1256DOS

LANDTEK SL-1256DOS Logging Personal Noise Dose Meter

LANDTEK SL-1256DOS

Art. number:
Date published 05.09.2026
Unavailable
Price and Delivery
Fill the request form below to receive a price and delivery offer

The LANDTEK SL-1256DOS is a personal noise dose meter with data logging capability, designed to evaluate workplace noise conditions by measuring accumulated noise exposure.

The instrument measures noise dose (DOSE) as a percentage of exposure relative to a selected criterion. For example, with a criterion of 90 dBA for 8 hours, 100% DOSE corresponds to 8 hours of exposure at 90 dBA. Other criteria can also be used to meet different national noise regulations.

The SL-1256DOS can simultaneously measure frequency-weighted noise exposure and peak sound levels. The instrument complies with GB/T 15952.

It includes 8 built-in standard dose measurement setups: OSHA80, OSHA90, MSHA-80, MSHA-90, DOD, ACGIH, ISO85, and ISO90, as well as 9 user-defined measurement setups.

The meter also provides a Sound Level Meter (SLM) mode, in which it can display sound level, LAeq, peak level, and SEL.

The instrument has two overlapping measurement ranges: 60–130 dB and 70–140 dB. It supports A and C RMS frequency weightings, Linear and C Peak weightings, as well as Fast and Slow time weightings.

Energy exchange rates of Q = 3, 4, 5, and 6 dB are supported, allowing the instrument to be configured for different occupational noise assessment standards and regulatory requirements.

In dose measurement mode, the display can show % DOSE, projected 8-hour dose, peak level, and measurement duration.

For applications based on European Union regulations or equivalent ISO standards, the instrument can measure the daily sound exposure level LAeq,8h, according to IEC 61252, equivalent to LEX,8h as required by ISO 1999. For OSHA/MSHA applications, it can measure TWA (Time Weighted Average).

The SL-1256DOS includes a built-in clock and automatic timing functions, allowing the user to preset measurement start time, stop time, and duration.

The logging interval can be set from 1 second to 1 hour, while the internal memory can store up to 120,000 data points.

The instrument can measure and store up to five specified Ln statistical noise levels.

It provides 50 storage locations for dose measurement results, and measurement setups can be downloaded from a PC to the meter.

A keypad lock function helps prevent accidental operation or changes to settings during measurement.

For data transfer and analysis, the instrument provides a USB data output for connection to a PC.

The LANDTEK SL-1256DOS is suitable for personal noise exposure assessment, occupational health and safety monitoring, industrial hygiene, workplace noise surveys, and professional noise monitoring in accordance with ISO, OSHA, MSHA, DOD, and ACGIH methodologies.

 

For more information:
Logging Personal Noise Dose Meter SL-1256DOS - Guangzhou Landtek Instruments Co.,LTD

Parameter Value
Applicable Standards IEC 61252, ANSI S1.25-1992 for noise dose meters and sound exposure meters; IEC 61672-1:2013 Class 2
Microphone 1/2" electret condenser microphone with 31" integral cable
Display LCD
Measurement Range 1 60–130 dB (A and C)
Measurement Range 2 70–140 dB (A and C)
Peak Range 1 93–133 dB Peak
Peak Range 2 103–143 dB Peak
Frequency Weighting – RMS Detector A or C
Frequency Weighting – Peak Detector C or Z (Linear)
Time Weighting Fast and Slow
Exchange Rate 3, 4, 5 or 6 dB
Stabilization Time 10 s after power-on
Threshold Level 70–90 dB, in 1 dB steps
Criterion Level 80, 84, 85 or 90 dB
Over-Range Indication 0.1 dB above the upper limit of the selected measurement range
High Level Detector 115 dB
Keypad Lock Lock: Up + Right arrows; Unlock: Left + Down arrows
Clock Real-time clock with calendar
Result Memory Up to 50 measurements
Preset Measurement Duration 5, 10, 15 or 30 min; 1, 2, 4, 8, 10, 12 or 24 h
Timers Up to 16 automatic start/stop timers, programmable up to 1 month in advance
Timer Control Via PC software or meter keypad
Data Logging All measurement parameter values can be logged
Logging Interval 1 s, 2 s, 5 s, 10 s, 15 s, 20 s, 30 s, 1 min, 2 min, 5 min, 10 min, 15 min, 20 min, 30 min or 60 min
Logging Capacity 120,000 values
Built-in Setups 8 predefined standard setups
User-defined Setups 9
Output USB interface
Power Supply 4 × 1.5 V AAA (UM-4) batteries
Battery Indicator 4-step battery level indication; flashes when voltage is insufficient
Operating Temperature 0 to +50 °C
Operating Humidity 10–90% RH
Storage Temperature -10 to +60 °C
Storage Humidity 10–75% RH
Dimensions 140 × 72 × 34 mm
Weight Approx. 350 g, including batteries
Standard Accessories Main unit, external probe, screwdriver, B04 carrying case, operation manual
Optional Accessories USB data interface/cable with software, Bluetooth data adapter with software

Preset Measurement Setups

Setup Range Time Weighting Frequency Weighting Peak Weighting Q Threshold Criterion User Adjustable
OSHA-80 70–140 dB Slow A Lin 5 80 dB 90 dB No
OSHA-90 70–140 dB Slow A Lin 5 90 dB 90 dB No
MSHA-80 70–140 dB Slow A Lin 5 80 dB 90 dB No
MSHA-90 70–140 dB Slow A Lin 5 90 dB 90 dB No
DOD 70–140 dB Slow A Lin 4 80 dB 85 dB No
ACGIH 70–140 dB Slow A Lin 3 80 dB 85 dB No
ISO-85 70–140 dB Fast A C 3 70 dB 85 dB No
ISO-90 70–140 dB Fast A C 3 70 dB 90 dB No
User 1–9 70–140 dB Fast A C 3 80 dB 90 dB Yes
SLM 70–140 dB Slow A Lin 3 N/A N/A Yes

Displayed Parameters by Mode

Depending on the selected setup, the instrument can display parameters such as DOSE%, PDOSE%, TWA, PTWA, LAVG, LEQ, SEL, MAX, PKZ, L10, L50, L90, L95, L99, SPL, LEPd, and SE (Pa²h). In SLM mode, the main displayed parameters are SPL, MAX, LEQ, and PKZ.