Senseca HD2022 – Class 2 Sound Calibrator for Sound Level Meters and Measurement Microphones
Senseca HD2022 – Class 2 Sound Calibrator for Sound Level Meters and Measurement Microphones
Senseca HD2022
Senseca HD2022 is a portable sound calibrator for checking and calibrating sound level meters and precision microphones in laboratory and field applications. It generates a 114.0 dB signal at 1000 Hz and complies with IEC 60942:2003, Class 2, and ANSI S1.40-1984.
The acoustic coupling cavity accepts standard ½″ microphones according to IEC 61094-1 and IEC 61094-4. The optional HD2020AD4 adapter enables use with ¼″ microphones.
A piezoceramic generator, internal sensor and automatic gain control maintain a stable sound level. The system minimises the influence of temperature, humidity and atmospheric pressure within the specified operating range. No static pressure correction is required between 65 and 108 kPa.
The instrument operates with one 9 V alkaline battery, type 6LR61, providing up to 48 hours of continuous operation. The LOW BATT LED warns of insufficient charge. If the voltage is too low, the instrument switches off or prevents operation.
The ABS housing provides IP64 protection, measures 53 × 43 × 83 mm and weighs 160 g. The operating temperature range is 0 to +40 °C at relative humidity up to 90%.
For more information on technical specifications, documentation and compatible accessories, visit the Senseca HD2022 page on the manufacturer’s website.
Technical specifications of Senseca HD2022
| Parameter | Value |
|---|---|
| Standards and compatibility | |
| Compliance | IEC 60942:2003, Class 2; ANSI S1.40-1984 |
| Coupling cavity | For standard ½″ microphones with a diameter of 12.7 ±0.03 mm, according to IEC 61094-1 and IEC 61094-4 |
| Optional adapter | HD2020AD4 for ¼″ microphones |
| Microphone equivalent volume | 5–250 mm³ |
| Acoustic characteristics | |
| Frequency | 1000 Hz |
| Frequency tolerance | 2% at 0 to +40 °C and 10–90% RH |
| Sound pressure level | 114.0 dB ±0.3 dB at 1 kHz, referenced to 101.3 kPa, 23 ±3 °C and 65% RH |
| Reference conditions | 23 °C, 50% RH, 101.3 kPa; microphone capsule with a 10 mm³ equivalent volume |
| Stabilisation time | 10 seconds |
| Total distortion | Less than 1% |
| Ambient influences | |
| Temperature and humidity influence | Less than 0.3 dB at 0 to +40 °C and 10–90% RH |
| Static pressure influence | Less than 0.1 dB at 65–108 kPa |
| Electromagnetic field effects | Less than 0.3 dB |
| Level stability | |
| Short-term stability | ±0.05 dB |
| Stability after one year of normal use | ±0.15 dB |
| Operating and storage conditions | |
| Operating temperature | 0 to +40 °C |
| Operating relative humidity | Up to 90% RH |
| Storage temperature | −25 to +70 °C |
| Power supply | |
| Battery | 9 V alkaline battery, IEC type 6LR61 |
| Alternative power | 9 V rechargeable batteries may also be used |
| Battery life | 48 hours of continuous operation with a good-quality alkaline battery |
| Housing and dimensions | |
| Housing material | ABS |
| Dimensions | 53 × 43 × 83 mm |
| Weight | 160 g |
| Ingress protection | IP64 |
| Ordering codes | |
| HD2022 | Acoustic calibrator supplied with a 9 V alkaline battery and user manual |
| HD2020AD4 | Adapter for ¼″ microphone capsules |