Senseca HD2024 – Class 1 Sound Calibrator for Sound Level Meters and Measurement Microphones
Senseca HD2024 – Class 1 Sound Calibrator for Sound Level Meters and Measurement Microphones
Senseca HD2024
Senseca HD2024 is a portable sound calibrator for checking and calibrating sound level meters and precision microphones in laboratory and field applications. It generates a stable sound pressure level of 94.0 dB at 1000 Hz, referenced to 20 µPa. It complies with IEC 60942:2003, Class 1, and ANSI S1.40-1984.
The acoustic coupling cavity accepts standard ½″ microphones according to IEC 61094-1 and IEC 61094-4. The optional HD2020AD4 adapter enables use with ¼″ microphones.
The piezoceramic generator and feedback system maintain a stable level under changing ambient conditions. Static pressure influence is less than 0.1 dB across the 65–108 kPa range.
The instrument operates with a 9 V alkaline battery, type 6LR61. The LOW BATT LED indicates insufficient charge. If the voltage is too low, the calibrator switches off or prevents operation. The ABS housing provides IP64 protection. The operating range is −10 to +50 °C at relative humidity up to 90%. The model was also developed with vehicle inspection applications in mind.
For more information on technical specifications, documentation and compatible accessories, visit the Senseca HD2024 page on the manufacturer’s website.
Technical specifications of Senseca HD2024
| Parameter | Value |
|---|---|
| Standards and compatibility | |
| Compliance | IEC 60942:2003, Class 1; ANSI S1.40-1984 |
| Coupling cavity | For standard ½″ microphones with a diameter of 12.7 ±0.03 mm, according to IEC 61094-1 and IEC 61094-4 |
| Optional adapter | HD2020AD4 for ¼″ microphones |
| Microphone equivalent volume | 5–250 mm³ |
| Acoustic characteristics | |
| Frequency | 1000 Hz |
| Frequency tolerance | 1% at −10 to +50 °C and 10–90% RH |
| Sound pressure level | 94.0 dB ±0.3 dB at 1 kHz, referenced to 101.3 kPa, 23 ±3 °C and 65% RH |
| Reference conditions | 23 °C, 50% RH, 101.3 kPa; microphone capsule with a 10 mm³ equivalent volume |
| Stabilisation time | 10 seconds |
| Total distortion | Less than 1% |
| Ambient influences | |
| Temperature and humidity influence | Less than 0.3 dB at −10 to +50 °C and 10–90% RH |
| Static pressure influence | Less than 0.1 dB at 65–108 kPa |
| Electromagnetic field effects | Less than 0.3 dB |
| Level stability | |
| Short-term stability | ±0.05 dB |
| Stability after one year of normal use | ±0.15 dB |
| Operating and storage conditions | |
| Operating temperature | −10 to +50 °C |
| Operating relative humidity | Up to 90% RH |
| Storage temperature | −25 to +70 °C |
| Power supply | |
| Battery | 9 V alkaline battery, IEC type 6LR61 |
| Alternative power | 9 V rechargeable batteries may also be used |
| Battery life | 48 hours of continuous operation with a good-quality alkaline battery |
| Housing and dimensions | |
| Housing material | ABS |
| Dimensions | 53 × 43 × 83 mm |
| Weight | 160 g |
| Ingress protection | IP64 |
| umn_prd_HD2024_en (1... | PDF, 560.8 KB | Download |