Senseca LPPHOT01 Photometric Probe for Indoor Illuminance Monitoring
Senseca LPPHOT01 Photometric Probe for Indoor Illuminance Monitoring
Senseca LPPHOT01
The Senseca LPPHOT01 is a photometric probe for illuminance measurement, designed for continuous indoor monitoring of lighting conditions. The probe measures illuminance in lux and features a spectral response matched to the standard photopic curve V(λ), providing measurements aligned with the sensitivity of human vision.
The LPPHOT01 is specified as a Class B luxmeter probe and uses a photodiode, optical filter, and diffuser. Incident light generates a photocurrent, which is converted into a potential difference across a shunt resistor. The resulting signal can be read directly with a suitable voltmeter or digital multimeter, with no external power supply required for the probe.
The measuring range is 0…200 klux, with a calibration accuracy of better than 4%. To improve angular response, the probe uses a cosine-corrected diffuser, with f₂ below 3%.
Each probe is individually calibrated, and its specific sensitivity factor is marked on the housing and used to calculate the measured value.
To achieve approximately 1% reading precision, the probe output should be connected to a digital multimeter with an input resistance greater than 100 kΩ.
The probe operates over a temperature range of 0…50 °C and is housed in a compact Ø30 mm × 38 mm enclosure with a 5 m cable for connection to the reading instrument.
For more information:
https://www.senseca.com/en/product/lpphot01-photometric-probe-for-indoor-illuminance-monitoring
Technical Specifications – Senseca LPPHOT01
| Parameter | Value |
|---|---|
| Typical Sensitivity | 0.5…1.5 mV/klux |
| Measuring Range | 0…200,000 lux |
| Spectral Response | V(λ) |
| Calibration Accuracy | <4% |
| f₁′ (V(λ) Match Error) | <6% |
| f₂ (Cosine Response / Directional Error) | <3% |
| f₃ (Linearity) | <1% |
| f₅ (Fatigue) | <0.5% |
| Operating Temperature | 0…50 °C |
| Output Impedance | 0.5…1 kΩ |
| Dimensions | Ø30 mm × 38 mm height |
The difference between the spectral response of the LPPHOT01 and the standard photopic curve V(λ) is evaluated by means of the f₁′ error.
Probe calibration is performed by comparison with a luxmeter calibrated by a Primary Metrological Institute, while the probe is illuminated with a standard illuminant A — a reference incandescent lamp with a colour temperature of 2856 K.
| dsh_fam_LPxx01_en (1... | PDF, 991.1 KB | Download |