Senseca PMsenseCR Particulate Matter and CO₂ Transmitter
Senseca PMsenseCR Particulate Matter and CO₂ Transmitter
Senseca PMsenseCR
The Senseca PMsenseCR is a high-precision particulate matter transmitter with an optional integrated CO₂ sensor, designed for environments where strict contamination control is essential, including pharmaceutical and high-tech cleanrooms.
The instrument uses laser scattering technology to detect and classify airborne particles according to their size. Its measuring concept is suitable for applications related to cleanroom contamination monitoring in accordance with ISO 14644 requirements.
PMsenseCR provides five particle-size measurement channels: Ch.1: >0.3 µm, Ch.2: >0.5 µm, Ch.3: >1.0 µm, Ch.4: >2.5 µm, and Ch.5: >5.0 µm.
The transmitter is designed for long-term stability, fast response, high sensitivity, and extended operating life, with no routine maintenance required.
An optional CO₂ sensor can be integrated on request, allowing simultaneous monitoring of airborne particles and carbon dioxide concentration in indoor environments.
The instrument can operate in either continuous or cyclic measurement mode. Cyclic operation helps extend sensor life, while continuous mode provides uninterrupted contamination monitoring.
For integration into automation and monitoring systems, the PMsenseCR supports an RS485 interface with MODBUS-RTU or proprietary ASCII protocol, as well as 4–20 mA analog outputs.
The Senseca PMsenseCR is suitable for cleanrooms, pharmaceutical manufacturing, high-tech production environments, laboratories, and other controlled areas where reliable monitoring of airborne particle concentration and particle size is required.
For more information:
PMsenseCR – PM and CO2 Transmitter - CO2 - Air Quality - Senseca
Technical Specifications – Senseca PMsenseCR
Particulate Matter
| Parameter | Value |
|---|---|
| Measuring Principle | Laser scattering |
| Measured Particles | Number of particles/m³ with size >0.3 µm; >0.5 µm; >1 µm; >2.5 µm; >5 µm |
| Measuring Range | <3.3 × 10⁹ particles/m³; loss of linearity above this threshold |
| Particle Size Detection Range | Ø 0.3–10 µm |
| Lowest Detectable Concentration | 350 particles/m³ |
| Detection Efficiency at 0.3 µm | 50% ±30% |
| Detection Efficiency at 0.5 µm | 100% ±20% |
| Detection Efficiency Standard | ISO 21501-4 |
| False Count | <3 over 15 min |
| Repeatability for Particles >0.3 µm and >0.5 µm | <3% |
| Repeatability for Particles >1 µm | <5% |
| Repeatability for Particles >2.5 µm | <10% |
| Repeatability for Particles >5 µm | <15% |
| Sensor Warm-Up Time | 15 s |
| Sensor Lifetime – Continuous Operating Mode | >10,000 hours at 1 measurement/s; default mode |
| Sensor Lifetime – Cyclic Operating Mode | Approx. 5 years with a 5-minute cycle |
CO₂ – PMBsenseCR… Models Only
| Parameter | Value |
|---|---|
| Measuring Principle | Dual-wavelength NDIR |
| Measuring Range | 0–5000 ppm |
| Accuracy | ±(50 ppm + 3% of measured value) at 25 °C and 1013 hPa |
| Response Time | <120 s at an air velocity of 2 m/s |
| Long-Term Stability | 5% of measured value / 5 years |
| Temperature Drift | 1 ppm/°C |
Outputs and Communication
| Parameter | Value |
|---|---|
| Digital Interface | RS485 with MODBUS-RTU or proprietary ASCII protocol |
| PMsenseCR-M / PMBsenseCR-M | RS485 MODBUS-RTU |
| PMsenseCR-A / PMBsenseCR-A | RS485 MODBUS-RTU + 2 analog outputs |
| Analog Current Outputs – Version A | 2 × 0/4–20 mA |
| Maximum Load Resistance for Current Output | 500 Ω |
| Optional Voltage Outputs | 2 × 0–10 V on request |
| Minimum Input Resistance for Voltage Output | 10 kΩ |
Power Supply and Consumption
| Parameter | Value |
|---|---|
| Power Supply | 7–30 V DC |
| Power Supply for 0–10 V Outputs | 15–30 V DC |
| Current Consumption During Measurement | 25 mA at 24 V DC |
| Current Consumption in Standby Mode | 4 mA, cyclic operating mode only |
| Consumption Note | The specified consumption does not include the current drawn by the analog outputs |
Construction and Operating Conditions
| Parameter | Value |
|---|---|
| Electrical Connection | M12 8-pole circular connector |
| Operating Temperature | -20 to +70 °C |
| Operating Relative Humidity | 0–95 %RH |
| Operating Atmospheric Pressure | 500–1500 hPa |
| Housing Material | Polycarbonate |
| Air Inlet Protection | Rain-proof and UV-resistant inlet air filter |
| Protection Rating | IP53 |
| Dimensions | 120 × 94 × 71 mm, excluding M12 connector |
| Weight | 330 g |
Ordering Code
| Code / Option | Description |
|---|---|
| PMsenseCR | Base model for particulate matter measurement |
| Blank | PM only |
| B | PM + CO₂ |
| M | RS485 MODBUS-RTU |
| A | RS485 MODBUS-RTU + 2 analog outputs |