Sonel PVM-1530 Max – Measurement and Diagnostic Kit for Photovoltaic Systems up to 1500 V

Sonel PVM-1530 Max

Sonel PVM-1530 Max – Measurement and Diagnostic Kit for Photovoltaic Systems up to 1500 V

Sonel PVM-1530 Max

Art. number:
Date published 07.11.2026
Delayed delivery
Price and Delivery
Fill the request form below to receive a price and delivery offer

The kit includes the PVM-1530 photovoltaic meter, two IRM-1 solar radiation and temperature meters, and the CMP-1015-PV digital clamp meter for photovoltaic systems.

Technical description

The Sonel PVM-1530 Max is an extended kit for testing, diagnosing and assessing the safety and performance of photovoltaic installations. The PVM-1530 supports Category 1 measurements under EN 62446-1 and Category 2 I‑V characteristic measurements under EN 62446-1 and EN 61829. It is suitable for standard, bifacial and high-efficiency PV modules.

The tester measures I‑V and P‑V characteristics and converts results to Standard Test Conditions (STC) in accordance with EN IEC 60891. The two IRM-1 meters measure solar irradiance, module temperature and ambient temperature. Their data helps assess measurement conditions, support PV system design and identify shading. If communication is temporarily interrupted, the reSYNC function automatically supplements the results with the parameters recorded by the IRM-1 meters once the connection is restored.

PVM-1530 measurement functions

  • Open-circuit voltage of a PV module or string: up to 1500 V DC
  • AC voltage: up to 1000 V, with frequency measurement
  • Short-circuit current of a PV module or string: up to 40 A DC
  • PV insulation resistance at test voltages of 250, 500, 1000 or 1500 V DC, with simultaneous RISO+ and RISO− measurements
  • AC circuit insulation resistance at 250, 500 or 1000 V DC
  • Protective conductor and equipotential bonding resistance using a ±200 mA current
  • PV operating current and AC current measurement using the CMP-1015-PV
  • AC/DC power measurement
  • Bypass diode testing with automatic polarity detection
  • Blocking diode testing at 1000 or 1500 V DC

IRM-1 and CMP-1015-PV functions

The IRM-1 measures solar irradiance in W/m² or BTU/ft²·h, PV module and ambient temperature, module tilt and orientation. Each IRM-1 includes a built-in compass, tilt sensor, LoRa communication and a data logger with memory for up to 5,000 records. Its enclosure is rated IP65.

The CMP-1015-PV measures DC voltage and current in photovoltaic installations up to 1500 V. Its VFD function supports measurements downstream of frequency converters. The meter also features a graphing data logger and a high-brightness display for outdoor use.

The PVM-1530 features touchscreen operation, a bright colour display, structured measurement memory, Bluetooth and Wi-Fi connectivity. Data can be transferred to a computer by cable or wirelessly. Sonel Reader enables users to view, export and print results. Sonel Reports Plus is available as optional software for preparing electric shock protection reports. The kit is part of the Sonel MeasureEffect™ platform.

For more information: Sonel PVM-1530 Max official product page

PVM-1530 | Technical specifications

Parameter Measuring range Display range Resolution Accuracy
Voltage        
AC voltage 0.0…1000.0 V 0.0…1000.0 V 0.1 V ±(2% of measured value + 6 digits)
DC voltage 0.0…1500.0 V 0.0…1500.0 V 0.1 V ±(0.5% of measured value + 2 digits)
Frequency 10.0…100.0 Hz 0…100.0 Hz 0.1 Hz ±(0.5% of measured value + 2 digits)
Short-circuit current ISC        
At 1500 V DC 0.00…30.00 A 0.00…30.00 A 0.01 A ±(1% of measured value + 2 digits)
At 1000 V DC 0.00…40.00 A 0.00…40.00 A 0.01 A ±(1% of measured value + 2 digits)
Insulation resistance        
AC-side insulation resistance        
Test voltage 250 V 250 kΩ…200 MΩ
according to EN IEC 61557-2
0.0 kΩ…200.0 MΩ from 0.1 kΩ ±(3% of measured value + 8 digits)
Test voltage 500 V 500 kΩ…500 MΩ
according to EN IEC 61557-2
0.0 kΩ…500.0 GΩ from 0.1 kΩ ±(3% of measured value + 8 digits)
Test voltage 1000 V 1000 kΩ…1.000 GΩ
according to EN IEC 61557-2
0.0 kΩ…1.000 GΩ from 0.1 kΩ ±(3% of measured value + 8 digits)
DC-side insulation resistance        
Test voltage 250 V / 500 V / 1000 V / 1500 V 250 kΩ…500 GΩ
according to EN IEC 61557-2
0.0 kΩ…500.0 GΩ from 0.1 kΩ ±(8% of measured value + 8 digits)
Resistance of protective conductors and equipotential bonding        
Measurement with ±200 mA current 0.10 Ω…1999 Ω
according to EN IEC 61557-4
0.00 Ω…1999 Ω from 0.01 Ω from ±(2% of measured value + 3 digits)
Operating current As for CMP-1015-PV As for CMP-1015-PV As for CMP-1015-PV As for CMP-1015-PV
Power measurement 0.0…999.0 kW 0.0…999.0 kW 0.1 kW ±(6% of measured value + 5 digits)

IRM-1 | Technical specifications

Parameter Measuring range Display range Resolution Accuracy
Solar irradiance        
Measurement in W/m² 0…1400 W/m² 0…1400 W/m² 1 W/m² ±(5% of measured value + 2 digits)
Measurement in BTU/ft²·h 0…444 BTU/ft²·h 0…444 BTU/ft²·h 1 BTU/ft²·h ±(5% of measured value + 2 digits)
PV module and ambient temperature        
Measurement in °C −20.0…100.0 °C −20.0…100.0 °C 0.1 °C ±(1% of measured value + 5 digits)
Measurement in °F −4.0…212.0 °F 4.0…212.0 °F 0.1 °F ±(1% of measured value + 5 digits)
Tilt angle −90°…+90° −90°…+90° 1° ±2°
Direction — compass 0°…360° 0°…360° 1° ±5°

Measured value — the value recorded by the instrument.

Note: In the supplied specification, the display range for insulation resistance at a 500 V AC test voltage is listed as 500.0 GΩ, while the measuring range is up to 500 MΩ. For temperature in °F, the measuring range starts at −4.0 °F, while the display range starts at 4.0 °F. Values are retained as stated in the source.

IRM-1 insobs v1.13 G... PDF, 1681 KB Download
PVM-1530 insobs v1.1... PDF, 1488.6 KB Download
datasheet_PVM-1530_E... PDF, 1406.8 KB Download